Model Number V60 |
Device Problem
Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 15oct2020.
|
|
Event Description
|
The customer reported the ventilator does not work.The unit was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or the user.
|
|
Manufacturer Narrative
|
G4:13jan2021.B4:14jan2021.The manufacturer's field service engineer (fse) confirmed the reported issue.The customer was provided with the part numbers for the power switch overlay and gas delivery subsystem (gds).After numerous attempts to obtain information on the repair of this device with no customer response, this complaint is being closed.If further information is obtained, a supplemental report will be submitted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4:30apr2021; b4:03may2021.The gas delivery system (gds) and data acquisition assembly (daq) were returned for failure analysis.Visual inspection on both parts revealed no anomalies.The customer complaint was verified.Root cause is failure of airflow sensor, caused by u1 drifting out of calibration.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|