• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 15oct2020.
 
Event Description
The customer reported the ventilator does not work.The unit was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or the user.
 
Manufacturer Narrative
G4:13jan2021.B4:14jan2021.The manufacturer's field service engineer (fse) confirmed the reported issue.The customer was provided with the part numbers for the power switch overlay and gas delivery subsystem (gds).After numerous attempts to obtain information on the repair of this device with no customer response, this complaint is being closed.If further information is obtained, a supplemental report will be submitted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4:30apr2021; b4:03may2021.The gas delivery system (gds) and data acquisition assembly (daq) were returned for failure analysis.Visual inspection on both parts revealed no anomalies.The customer complaint was verified.Root cause is failure of airflow sensor, caused by u1 drifting out of calibration.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10687620
MDR Text Key213682452
Report Number2031642-2020-03744
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/25/2020
Initial Date FDA Received10/15/2020
Supplement Dates Manufacturer Received09/25/2020
09/25/2020
Supplement Dates FDA Received01/14/2021
05/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-