• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number 20226
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/18/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent dislodgement occurred.A 8.0x40x135 cm express ld stent was opened during preparation.However, the stent slid off the balloon and dislodged from the catheter when removing the protective cover over the stent.The procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that stent dislodgement occurred.A 8.0x40x135 cm express ld stent was opened during preparation.However, the stent slid off the balloon and dislodged from the catheter when removing the protective cover over the stent.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: an express ld device was returned for analysis.The stent was not received.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A visual and microscopic examination identified no issues with the balloon material that could have contributed to the complaint incident.A visual and microscopic examination identified no issues with the tip of the device during analysis that could have contributed to the complaint incident.A visual and tactile examination identified no damage or any issues along the length of the device.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10689885
Report Number2134265-2020-14318
Device Sequence Number16881448
Product Code NIO
UDI-Device Identifier08714729392477
UDI-Public08714729392477
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/03/2023
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0025296641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Initial Date Received by Manufacturer 10/05/2020
Supplement Date Received by Manufacturer12/07/2020
Initial Report FDA Received Date10/16/2020
Supplement Report FDA Received Date12/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-