| Model Number |
482618570 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Doctor has kept the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the tulip of a xia serrato polyaxial screw located at s5 "came apart" from the shaft post-operatively.Revision surgery has been performed where the screw was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Photos were provided showing the tulip disengaged from the shank.Photos were not clear enough to make out any deformation or witness marks on the screw.A review of the device and complaint history records could not be performed as a valid lot code was not provided and could not be obtained.As the device was not returned and no further information was provided, a cause cannot be determined.According to the ifu, possible causes include patient activity level, patient fall, patient weight, and/or improper implantation.
|
| |
|
Event or Problem Description
|
|
It was reported that the tulip of a xia serrato polyaxial screw located at s5 "came apart" from the shaft post-operatively.Revision surgery has been performed where the screw was explanted.
|
| |
|
Search Alerts/Recalls
|