| Model Number |
COMPRESSION SCREW, Ø2.7X32MM |
| Medical Device Problem Codes |
Fracture (1260); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Nonunion of the bone occurred most likely due to the osteopenic condition of the patient's bone.It appears that the cyclic loading after the nonunion resulted in the unscrewing of the connector and the failure of the most distal cortical screw.Both consequences are predictable considering the long period of implantation with the nonunion (more than 2 years).
|
| |
|
Event or Problem Description
|
|
A compression screws attaching the plate to the bone broke being implanted in an infant patient.Initial surgery: (b)(6) 2018 (b)(6)of age).2 years and 3 months implantation before detection of failure.
|
| |
|