• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Date of Event 09/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluation: product testing could not be performed because the product was not returned as the iol was disposed.A product quality deficiency could not be confirmed.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search revealed that no other complaints for this production order number have been received.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.Note: the device was manufactured at the kulim site which has been closed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported intraocular lens (iol) was implanted in the patient¿s ocular sinister (left eye).The iol was explanted due to complaints of lens rotation.The incision was enlarged.There was no suture(s) and no vitrectomy.The lens was disposed and is not being returned.Through follow-up, additional information was received confirming the lens rotated 13 degrees and patient reported blurry vision.A jjsv lens with different model and diopter size was implanted.Patient outcome was reported as vision doing ok adjusting with some irritation.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key10694670
Report Number2020664-2020-00052
Device Sequence Number11556926
Product Code HQL
UDI-Device Identifier05050474553361
UDI-Public(01)05050474553361(17)211019
Combination Product (Y/N)Y
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2021
Device Model NumberZCT300
Device Catalogue NumberZCT300U220
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2020
Initial Report FDA Received Date10/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-