• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VIDEOARTHORSCOPE HD 4MM X 70 DEG; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VIDEOARTHORSCOPE HD 4MM X 70 DEG; ARTHROSCOPE Back to Search Results
Model Number 72202962S
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is currently under investigation by the manufacturing registered site.
 
Event or Problem Description
It was reported that during hip arthroscopy procedure, the scope was scratched.No significant delay and a back up was available to complete the procedure.No other complications were reported.Preliminary results of investigation have concluded that the scope had a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3,h6: the device, which was used in a procedure, was returned for evaluation.There was a relationship between the reported event and the device.A visual inspection was performed and showed the scope to have distal tip and fiber damage and a cracked and scratched distal lens.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.This failure is confirmed not originate from manufacturing, packaging, or labeling defects.This damage is caused by contact with another source.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.No manufacturing related defects were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOARTHORSCOPE HD 4MM X 70 DEG
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10694675
Report Number3003604053-2020-00128
Device Sequence Number3933240
Product Code HRX
UDI-Device Identifier00885554024210
UDI-Public00885554024210
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K043395
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202962S
Device Catalogue Number72202962S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer10/29/2020
Initial Report FDA Received Date10/17/2020
Supplement Report FDA Received Date10/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-