Model Number 72203967S
Device Problems No Display/Image (1183); Electrical Shorting (2926); Intermittent Energy Output (4025)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2020
Event Type  malfunction  
Event Description
It was reported that during shoulder scope procedure, the camera head seemed to be shorting out, it went from working fine to flickering off and on.No delay and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of electrical short and flickering could not be reproduced.Product passed functional testing and 2 hour burn-in.Electrical shorting and flickering did not occur during functional testing and burn-in.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10694696
MDR Text Key211925808
Report Number1643264-2020-01546
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (y/n)N
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Date Manufacturer Received11/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-