• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 09/25/2020
Event Type  Malfunction  
Event Description

It was reported that the motor drive unit was not being recognized. No case reported. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10694701
MDR Text Key211893954
Report Number1643264-2020-01548
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 11/13/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/17/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number72200616SR
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer10/06/2020
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/12/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-