Catalog Number 72200616SR |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/25/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the motor drive unit was not being recognized.No case reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of recognition failure could not be reproduced.Product failed for motor stall error but test control unit recognized handpiece.Recognition failure did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|