Additional Manufacturer Narrative
|
|
The subject device referenced in this report was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
| |
|
Event or Problem Description
|
|
During an unspecified procedure, the subject device was used.It was reported that the plastic protection began to melt.There was no patient injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The tissue pad of the subject device was worn and partially separated.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the tissue pad was damaged since the user continued to activate output without grasping tissue (including after the tissue already cut).The above device handling has warned in the instruction manual.
|
| |
|