• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE ONE STEP BUTTON; TUBE, GASTRO-ENTEROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE ONE STEP BUTTON; TUBE, GASTRO-ENTEROSTOMY Back to Search Results
Model Number M00568520
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter address: (b)(6).(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive one step button was used during a percutaneous endoscopic gastrostomy (peg) placement procedure (b)(6) 2020.According to the complainant, during the procedure, as the one step button was passed through the patient, the catheter was torn and separated.The procedure was completed with a new endovive one step button.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block e1 (initial reporter address 1): (b)(6).Block h6 (device codes): problem code 2907 captures the reportable event of one step button delivery system detached.H10: a visual examination of the device revealed that the button was not released.The c-flex tubing was broken, resulting in button delivery system separation.Therefore, the reported complaint was confirmed.It was noted the c-flex tubing was narrowed near the broken area indicating stretching of material due to tensile force and the broken area was irregular.The outer diameter of the c-flex tubing was measured in some points of the device, the outer diameter was found out of specification in the narrow section confirming that the component was stretched.Probably the c-flex tubing was broken/separated due to user technique, patient anatomy or other procedural factors such as the incision size, also force applied to pull the device during placement could have contributed to the event, stretching material found on the c-flex tubing suggests that the component was submitted to a tensile force.Based on the information available and the analysis performed, the investigation conclusion code for the complaint will be documented as adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive one step button was used during a percutaneous endoscopic gastrostomy (peg) placement procedure (b)(6) 2020.According to the complainant, during the procedure, as the one step button was passed through the patient the catheter was torn and separated.The procedure was completed with a new endovive one step button.There were no patient complications reported as a result of this event.Additional information received on october 14, 2020.One step button tube got broken near the button.Reportedly, the detached portion was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE ONE STEP BUTTON
Common Device Name
TUBE, GASTRO-ENTEROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10695557
Report Number3005099803-2020-04727
Device Sequence Number3708807
Product Code KGC
UDI-Device Identifier08714729149231
UDI-Public08714729149231
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2023
Device Model NumberM00568520
Device Catalogue Number6852
Device Lot Number0025568201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer10/14/2020
Initial Report FDA Received Date10/19/2020
Supplement Report FDA Received Date11/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-