• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS ADVANCE XP SLING SYSTEM; MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS ADVANCE XP SLING SYSTEM; MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE Back to Search Results
Model Number 720163-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 09/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product investigation completed.As the complaint component was not returned for analysis, no product analysis could be performed.The reported allegations could not be confirmed.The event cannot be reproduced or substantiated; therefore, no escalation to ncep, capa or scar is required.
 
Event or Problem Description
It was reported that during a sling implant procedure the first helical needle was successfully passed and connected to the sling.The second helical needle perforated the urethra and was connected to the device.When the sling was pulled through, the urethral injury was noticed.The physician removed the entire device and repaired the perforation.The procedure was then aborted due to this injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ADVANCE XP SLING SYSTEM
Common Device Name
MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
100 boston scientific way
marlborough MA 01752 1242
MDR Report Key10695581
Report Number2183959-2020-04682
Device Sequence Number11583450
Product Code OTM
UDI-Device Identifier00878953005560
UDI-Public00878953005560
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model Number720163-01
Device Catalogue Number720163-01
Device Lot Number0025892496
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/25/2020
Initial Report FDA Received Date10/19/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
-
-