• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION Back to Search Results
Model Number 71340007
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/28/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery, the 2.0 accord cable frayed while inside the patient.Item was retrieved from patient and procedure concluded with a delay of 5 minutes and without an injury.It is unknown if a backup device from smith and nephew was used to complete the procedure.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A few strands of wire fractured on the device, causing it to fray and become inoperable.The clinical/ medical evaluation concluded that this case reports the accord cable frayed while inside the patient during use.Per email communication, there was no patient injury, and the procedure was completed with a minimal delay using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.The potential probable cause of this event is likely overloading.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACC 2.0MM COCR CABLE W/CLAMP
Common Device Name
CERCLAGE, FIXATION
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10696478
Report Number1020279-2020-05487
Device Sequence Number3058350
Product Code JDQ
UDI-Device Identifier03596010485731
UDI-Public03596010485731
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K031162
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71340007
Device Catalogue Number71340007
Device Lot Number18KSM0286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer10/01/2021
Initial Report FDA Received Date10/19/2020
Supplement Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-