| Model Number |
ASD44A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
09/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Limited imaging was provided for evaluation.Images show what appears to be an echogenic mass that is consistent with the appearance of thrombus on the right atrial disc of a gore® cardioform asd occluder.The gore® cardioform asd occluder instructions for use list device thrombosis or thromboembolic event resulting in clinical sequelae as potential clinical and device adverse events.
|
| |
|
Event or Problem Description
|
|
It was reported the physician implanted a 44mm gore® cardioform asd occluder to treat an atrial septal defect on (b)(6) 2020.At the patient's one month follow-up, imaging revealed thrombus on the right atrial disc.The patient has been prescribed blood thinners.
|
| |
|