It was reported that during a procedure no locking screw was inserted in one of the distal locking holes.Once reduction was retained, the cortex screw was removed and a locking screw was inserted.The locking tabs, however, were too damaged and no screw could be placed in that hole.A delay of less than or equal to 30 minutes was reported.The procedure was completed with the same device.
|
|
H6, the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, we are unable to determine if adequate fixation was achieved without the placement of the distal screw.However, it was reported the procedure completed with the same device.According to the report, a delay of less than or equal to 30 minutes was reported.Since no harm has been alleged to this patient, no further clinical/medical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed device.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|