MAUDE Adverse Event Report: INSTITUT STRAUMANN AG SP IMPL., 3.3MM RN, SLACTIVE 14MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG SP IMPL., 3.3MM RN, SLACTIVE 14MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number SLACTIVE TITANIUM TL
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: production / design flaws, inadequate fitting of the suprastructure, occlusal overload, bruxism, bone resorption around the implants, insufficient axle alignment of the implants, trauma, biomechanical causes, design of the suprastructure.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2020 in fdi 23.On (b)(6) 2020, fracture of the implant was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: a) production / design flaws, b) inadequate fitting of the suprastructure, c) occlusal overload, d) bruxism, e) bone resorption around the implants, f) insufficient axle alignment of the implants, g) trauma, h) biomechanical causes and i) design of the suprastructure.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2020 in fdi 23.On (b)(6) 2020, fracture of the implant was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SP IMPL., 3.3MM RN, SLACTIVE 14MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
CH  04002
MDR Report Key10700176
Report Number0009613348-2020-59303
Device Sequence Number19895170
Product Code DZE
UDI-Device Identifier07630031714848
UDI-Public07630031714848
Combination Product (Y/N)N
Initial Reporter CountryDE
PMA/510(K) Number
K033984
Number of Events Summarized1
Summary Report (Y/N)Y
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2014
Device Model NumberSLACTIVE TITANIUM TL
Device Catalogue Number033.154S
Device Lot NumberW7425
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2020
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/16/2020
Event Location Other
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/19/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-