MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBE-150
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Codes Perforation (2001); Device Embedded In Tissue or Plaque (3165)
Date of Event 09/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Return of the device for analysis is anticipated.If the device is received for analysis, a supplemental report will be submitted when the device analysis is completed.Csi id: (b)(4).
 
Event or Problem Description
A stealth peripheral orbital atherectomy device (oad) was used to treat a chronic total occlusion in the mid and proximal anterior tibial artery (at).During treatment, an audible increase in pitch was noted.The oad was stopped and an attempt was made to remove the device, however the crown was stuck in the vessel.Several attempts were required to remove the device via pulling, and an angiogram following removal showed a small perforation.Multiple balloon inflations were performed and heparin was administered over a 25-30 minute period to resolve the perforation.The tip of the oad appeared damaged following removal.The patient was discharged and was monitored over the following 72 hours.There were no additional complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
st. paul, mn MN 55112
MDR Report Key10701575
Report Number3004742232-2020-00324
Device Sequence Number11745822
Product Code MCW
UDI-Device Identifier10850000491332
UDI-Public(01)10850000491332(17)220331(10)319814
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberDBE-150
Device Catalogue Number7-10059-08
Device Lot Number319814
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2020
Initial Report FDA Received Date10/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
-
-