• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG LINOX SMART T 65; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG LINOX SMART T 65; ICD LEAD Back to Search Results
Model Number 369820
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
This lead was explanted and replaced due to fracture.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
Upon receipt, the lead under complaint was found capped 11cm distal to the is1 connector pin.Only the proximal fragment was received for analysis.The returned lead fragment was visually analyzed.The visual inspection revealed that the insulation was, apart from the present fragmentation, free of breaches.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported clinical observation.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINOX SMART T 65
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key10701627
Report Number1028232-2020-04500
Device Sequence Number19626702
Product Code NVY
UDI-Device Identifier04035479112998
UDI-Public04035479112998
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2014
Device Model Number369820
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Initial Date Received by Manufacturer 10/16/2020
Supplement Date Received by Manufacturer10/20/2020
05/11/2021
Initial Report FDA Received Date10/19/2020
Supplement Report FDA Received Date10/26/2020
05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age35 YR
-
-