The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is severely bent and has signs of damage from implantation / extraction.The clinical medical investigation concluded that the responses to clinical requests were not provided, therefore, s+n has not received the adequate documentation to fully evaluate the root cause of the reported pain nor breakages.The four undated/unlabeled photos confirmed the reported breakages.Based on the information provided another tan nail and screw were implanted.The patient impact beyond the reported event could not be determined based on the limited information provided.No further medical assessment is warranted at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Some potential probable causes for this event could include surgical technique or a traumatic injury.Based on this investigation, the need for corrective action is indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|