MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 9.0MM ENDCUTTING IRH; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 9.0MM ENDCUTTING IRH; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71118231
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the edge of the blades broke after the reamer hit the k-wire.The surgeon used a 2.4 mm k-wire to block during reaming.No delay reported.No patient injuries reported.An s&n backup was available.
 
Additional Manufacturer Narrative
The device, used in treatment, were returned for evaluation.A visual inspection confirmed the pilot nose reamer has nicks and burrs.Also the pilot nose reamer is dull.This device shows significant signs of wear/usage.The date of manufacture is unknown for this device.A medical investigation was conducted and this case reports that the edge of the blades broke after the reamer hit the k-wire.Per complaint details, all the pieces were removed, and there was no injury or delay.The procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The device, used in treatment, were returned for evaluation.A visual inspection confirmed the pilot nose reamer has nicks and burrs.Also the pilot nose reamer is dull.This device shows significant signs of wear/usage.The date of manufacture is unknown for this device.A medical investigation was conducted and this case reports that the edge of the blades broke after the reamer hit the k-wire.Per complaint details, all the pieces were removed, and there was no injury or delay.The procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9.0MM ENDCUTTING IRH
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10702636
Report Number1020279-2020-05559
Device Sequence Number3630297
Product Code JDG
UDI-Device Identifier03596010416339
UDI-Public03596010416339
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71118231
Device Catalogue Number71118231
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/24/2020
Supplement Date Received by Manufacturer11/16/2020
11/16/2020
Initial Report FDA Received Date10/19/2020
Supplement Report FDA Received Date12/09/2020
12/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-