Model Number 86-6401
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); No Code Available (3191)
Date of Event 09/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed via tka on (b)(6) 2011.It was reported that the revision surgery was performed on (b)(6) 2020.The surgeon found that the screw of the explanted femoral stem¿s part loosened and very wore.No further information is available.
 
Event or Problem Description
Additional information received indicated that the reason for surgery was the sinking of the tibial side and then occurring the fracture.Bone fracture was confirmed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the received device was forwarded to commercialized product development for evaluation.No evidence was found of product malfunction or product error as a contributing factor to the reported event and the need for corrective action is not indicated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: b5, d7a, d9, d10 corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGTIBCEMSTM 13X30 1.5/2/2.5/3
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10705312
Report Number1818910-2020-22841
Device Sequence Number3025836
Product Code JWH
UDI-Device Identifier10603295216896
UDI-Public10603295216896
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K882234
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number86-6401
Device Catalogue Number866401
Device Lot NumberE75FD4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2020
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer02/03/2021
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date02/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA STAB INS S1.5 15MM.; PFC* MOD TIB TRAY NP, SZ 1.5.; PFC*SIGMA C/S NPOR FE RT SZ1.5.; PFC*SIGMA DIS AUG 8MM/1.5,RGHT.; PFC*SIGMA DIS AUG 8MM/1.5,RGHT.; PFC*SIGMA POST AUG 4MM SZ1.5.; PFC*SIGMA POST AUG 4MM SZ1.5.; PFCSIG +/-2FEMOFFSET STMBLT.; PFCSIGM FEMFLUTROD12X125MM7DEG.
Outcome Attributed to Adverse Event Required Intervention;
-
-