• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC* MOD TIB TRAY NP, SZ 1.5; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC* MOD TIB TRAY NP, SZ 1.5; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Model Number 86-6027
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Bone Fracture(s) (1870); No Code Available (3191)
Date of Event 09/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed via tka on (b)(6) 2011.It was reported that the revision surgery was performed on (b)(6) 2020.The surgeon found that the screw of the explanted femoral stem¿s part loosened and very wore.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the received device was forwarded to commercialized product development for evaluation.Review of the received device revealed the device was damaged.The investigation found no evidence of product malfunction or product error and the need for corrective action was not established.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: b5 and d10 corrected: h3.
 
Event or Problem Description
Additional information received indicated that the reason for surgery was the sinking of the tibial side and then occurring the fracture.Bone fracture was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10705313
Report Number1818910-2020-22840
Device Sequence Number16900496
Product Code JWH
UDI-Device Identifier10603295216827
UDI-Public10603295216827
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K961685
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-6027
Device Catalogue Number866027
Device Lot Number3233369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2020
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer11/08/2020
02/03/2021
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date11/19/2020
02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA STAB INS S1.5 15MM.; PFC*SIGMA C/S NPOR FE RT SZ1.5.; PFC*SIGMA DIS AUG 8MM/1.5,RGHT.; PFC*SIGMA DIS AUG 8MM/1.5,RGHT.; PFC*SIGMA POST AUG 4MM SZ1.5.; PFC*SIGMA POST AUG 4MM SZ1.5.; PFCSIG +/-2FEMOFFSET STMBLT.; PFCSIGM FEMFLUTROD12X125MM7DEG.; SIGTIBCEMSTM 13X30 1.5/2/2.5/3.
Outcome Attributed to Adverse Event Required Intervention;
-
-