BD CARIBE LTD. NEEDLE SPINAL S/SU 22GA 3IN QUINCKE; ANESTHESIA CONDUCTION NEEDLE Back to Search Results
Model Number 405171
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Pma/510(k)#: preamendment a device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported that the needle spinal s/su 22ga 3in quincke experienced incorrect label information and a mix of product types in a pack.Product defect was noted prior to use.The following information was provided by the initial reporter: material no.405171, batch no.9290118.Wrong needles inside the box.Received the needles 405160, lot # 8080923 in a box marked 405171.Qty: 1 box.
 
Event or Problem Description
It was reported that the needle spinal s/su 22ga 3in quincke experienced incorrect label information and a mix of product types in a pack.Product defect was noted prior to use.The following information was provided by the initial reporter: material no.405171; batch no.9290118 wrong needles inside the box.Received the needles 405160, lot # 8080923 in a box marked 405171.Qty: 1 box.
 
Additional Manufacturer Narrative
H6: investigation summary: two pictures were received by our quality team for evaluation.The photos showed two descriptions of spinal needles, one is the description of the shelf carton that states a material no.405171 (needle spinal s/su 22ga 3in quincke) batch no.9290118 with an expiration date of 30sep2024 and then the needle (individual blister) that is material no.405160 (needle sp s/su 20ga 1-1/4in introducer) lot 8080923 with an expiration date of 31mar2023.Based on the lot numbering material 405171 (22ga 3in quincke) batch no.9290118 is manufactured on oct2019 and the other material 405160 (20ga 1-1/4in introducer) lot 8080923 lot was manufactured a year earlier on apr2018.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.After the investigation, juncos controls regarding reconciliation and label generation was found adequate and did not show a possible root cause for the customer indicated condition.For each lot manufactured and packed at the site, a line clearance is completed and reconciliation is generated.Thus, bd juncos does not retain packed samples after the packaging process was completed.No possible cause has been identified for this complaint related to juncos manufacturing process within this investigation.A possible cause could be that the mix was generated after the bd process.Both lots were received by the same customer.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE SPINAL S/SU 22GA 3IN QUINCKE
Common Device Name
ANESTHESIA CONDUCTION NEEDLE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key10706000
Report Number2618282-2020-00083
Device Sequence Number3486371
Product Code BSP
UDI-Device Identifier30382904051717
UDI-Public30382904051717
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
SEE H10
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2024
Device Model Number405171
Device Catalogue Number405171
Device Lot Number9290118
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2020
Supplement Date Received by Manufacturer09/25/2020
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date10/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-