U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC MESH CUTTER; INSTRUMENT, CUTTING, ORTHOPEDIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC MESH CUTTER; INSTRUMENT, CUTTING, ORTHOPEDIC Back to Search Results
Model Number 391.952
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a synthes employee.Part number: 391.952, lot number: t936211, manufacturing site: (b)(4), release to warehouse date: 17-jul-2009.A review of the device history records was performed for the finished device lot number, and no non-conformance's were identified.Visual inspection: the mesh cutter (p/n: 391.952, lot #: t936211) was returned and received at us customer quality (cq).Upon visual inspection, the cutting jaw was observed to be broken and the broken fragment was not returned.There were scratches and discoloration was observed on the device but have no impact on the device functionality.No other issues were identified with the returned device.Device failure/defect identified? yes.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Service and repair evaluation the customer reported the device will not cut.The repair technician reported the distal tip was broken.Cutting jaws broken is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Document/specification review based on the date of manufacture, the current and manufactured revision of drawings were reviewed: mesh plate cutter.Complaint confirmed? yes, the device received was broken.Hence confirming the allegation.Investigation conclusion the complaint condition is confirmed for the mesh cutter (p/n: 391.952, lot #: t936211).There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, an inventory check occurred and multiple devices including a mesh cutter, right angle bender, plate cutter, 90 degree screwdriver t-handle, handle for 90° screwdriver, shaft for 90° screwdriver, and matrixrib screwdriver blade self-retaining f/90° screwdriver were found with malfunctions.When tested, the devices would not work properly.They would not turn, hold screws, and some were stripped.Some instruments were bent or dull.There was no patient or procedure involvement.This report is for one (1) mesh cutter.This is report 9 of 9 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESH CUTTER
Common Device Name
INSTRUMENT, CUTTING, ORTHOPEDIC
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10706281
Report Number2939274-2020-04739
Device Sequence Number11695705
Product Code HTZ
UDI-Device Identifier10887587041001
UDI-Public10887587041001
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number391.952
Device Catalogue Number391.952
Device Lot NumberT936211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Initial Date Received by Manufacturer 10/15/2020
Supplement Date Received by Manufacturer10/23/2020
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date10/23/2020
Patient Sequence Number1
-
-