| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cataract (1766); Capsular Bag Tear (2639)
|
| Date of Event |
09/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lens was returned in liquid in lens vial.Visual inspection found the haptics torn.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon inserted and removed a 13.2mm tmicl13.2 implantable collamer lens, -13.00/+3.5/094 (sphere/cylinder/axis), during the same surgery, due to an anterior subcapsular opacity and a capsular tear had occurred.Cataract surgery was performed and an iol lens was implanted in the ciliary sulcus.The reporter indicated the cause of the event was due to 2 pits were noted in the lens capsule (anterior) directly below the iridotomies.When the lens was inserted a capsular tear extended from the pits.
|
| |
|