• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, Back to Search Results
Model Number 71365705
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 09/22/2020
Event Type  malfunction  
Event Description
It was reported that during inspection the metal plate at the bottom of the anthology inserter poster hard, that sits inside of the exterior metal framework, was totally snapped off.No case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.(b)(4).
 
Manufacturer Narrative
D4: lot information updated.
 
Manufacturer Narrative
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The end of the device is completely broken off, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER POSTER HARD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10707740
MDR Text Key212177865
Report Number1020279-2020-05572
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010555663
UDI-Public03596010555663
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365705
Device Catalogue Number71365705
Device Lot Number18HM08532
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2021
Date Manufacturer Received09/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-