V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 20oct2020.
 
Event Description
The customer reported error blower temperature high.There was no patient involvement.
 
Manufacturer Narrative
G4:04nov2021.B4:03mar2021.H11:g5:k102985.H10: the manufacturer's field support engineer (fse) evaluated the unit and confirmed the reported error code in the diagnostic log.Air inlet was found to be clean and not blocked.Fse confirmed fan was operational.Fse recommended blower replacement due to error code.The fse replaced the unit's blower to resolve the reported issue.Unit passed required performance verification tests per philips standards.G4:06feb2021.B4:03mar2021.The blower was returned to the manufacturer for failure investigation.After testing, the customer's reported complaint could not be verified.The returned blower passed all testing, and no-fault was found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10708393
MDR Text Key213682857
Report Number2031642-2020-03782
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2020
Initial Date FDA Received10/20/2020
Supplement Dates Manufacturer Received09/29/2020
Supplement Dates FDA Received03/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse