SMITH & NEPHEW, INC. UNIVERSAL CHUCK; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71170183 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the universal chuck seems to be broken.It is unknown where the event occurs.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the universal chuck tip has multiple scratches, gouges and burrs in the metal.The locking mechanism is also damaged on the device causing the stated failure.The device shows significant signs of wear/usage.The device was manufactured in 2016.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|