• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012270-30
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354); Material Rupture (1546); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.The reported inflation problem was able to be confirmed.Additionally returned goods analysis identified there was fluid was leaking from the tip at 16 atmospheres.There was a pinhole in the proximal taper of the balloon and a tear noted in the inner member.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no report of a leak during preparation for use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the guide wire and/or other devices resulted in the noted inner member tear, the noted scratched proximal balloon marker and the noted balloon material pinhole; thus resulting in the reported inflation problem/noted leaks (pinhole, tip).The noted multiple bends on the hypotube likely occurred due to handling or during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that a 2.0x30mm mini trek rx balloon dilatation catheter failed to inflate.The procedure was successfully completed with an unspecified balloon catheter.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.Returned goods analysis identified there was fluid was leaking from the tip at 16 atmospheres.There was a pinhole in the proximal taper of the balloon and a tear noted in the inner member.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10708737
Report Number2024168-2020-08761
Device Sequence Number17036690
Product Code LOX
UDI-Device Identifier08717648138188
UDI-Public08717648138188
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number1012270-30
Device Catalogue Number1012270-30
Device Lot Number81212G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2020
Initial Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-