Model Number ZCT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Therapeutic Effects, Unexpected (2099); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown/ not provided.Date of event: unknown, not provided.Best estimate was (b)(6) 2020 to (b)(6) 2020.If explanted, give date: unknown/not provided.(b)(4).Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a doctor had to exchange one toric intraocular lens for another.This was due to lens rotated, and it was replaced with a model zcu300 22.5 diopter lens, because he was slightly plus.It was also noted that the exchange went well with no patient complications.
 
Additional Manufacturer Narrative
Device evaluation: product evaluation was not performed because the product has not been returned.If product is received after initial closure, the investigation request (ir) and complaint file (if necessary) may be re-opened to complete the return investigation.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10708858
Report Number9614546-2020-00451
Device Sequence Number17006903
Product Code HQL
UDI-Device Identifier05050474553354
UDI-Public(01)05050474553354(17)220329
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2022
Device Model NumberZCT300
Device Catalogue NumberZCT300U215
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer11/16/2020
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date11/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-