• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335562
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a thr procedure the reamer handle was not reaming correctly.Surgeon thought that this caused the cup to not sit properly.No injuries or surgical delays reported.Surgery was finished with the same device.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The failure mode pertains with the reamer handle.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.The probable cause for this event is likely due to a broken instrument.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 62MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10708950
Report Number1020279-2020-05597
Device Sequence Number3733084
Product Code MBL
UDI-Device Identifier03596010598295
UDI-Public03596010598295
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2025
Device Model Number71335562
Device Catalogue Number71335562
Device Lot Number15MM13199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer08/04/2021
Initial Report FDA Received Date10/20/2020
Supplement Report FDA Received Date08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-