SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Model Number 71335562 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/23/2020 |
Event Type
Injury
|
Event Description
|
It was reported that during a thr procedure the reamer handle was not reaming correctly.Surgeon thought that this caused the cup to not sit properly.No injuries or surgical delays reported.Surgery was finished with the same device.
|
|
Manufacturer Narrative
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The failure mode pertains with the reamer handle.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.The probable cause for this event is likely due to a broken instrument.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|