• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10013
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/30/2020
Event Type  malfunction  
Event Description
It was reported that, during calibration in a cori tka procedure, they had to insert the burr at least 20 times in order to lock it in successfully as when he pulled gently it kept unlocking.They tried both long attachments and it happened for each.There was a delay of less than 30 minutes.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill, part number rob10013, serial unknown and used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation, or field service report.A complaint history review for similar reported/confirmed complaints has identified prior events.While the reported problem was not confirmed during evaluation, the most probable contributing factor(s) to the reported problem is damage to the long attachment region that engages with the drill as a result of wear and tear.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10710894
MDR Text Key212278642
Report Number3010266064-2020-01915
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757321
UDI-Public00885556757321
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? No
Date Manufacturer Received10/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-