| Model Number |
401723M |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) associated with lot 17908009 revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported event.Additional information is pending and will be sent in upon 30 days after receipt.This is one of two devices involved in the reported event.The reporting reference number of the related device is 9616099-2020-04001.
|
| |
|
Event or Problem Description
|
|
As reported, the valve of the two (2) 7f 23cm brite tip catheter sheath introducer (csi) leaked a lot and the cap coming off the sheath just by taking out the unknown wire from the sheath.The physician was able to take out the leaking sheath and replace it with a new unknown sheath.There was no reported patient injury.The operator was trained to the brite tip sheath.The devices were opened in sterile filed.Only one device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
As reported, the valve of the two (2) 7f 23cm brite tip catheter sheath introducers (csi) were leaking and the cap came off the sheath when the user took out the unknown wire from the sheath.The physician was able to take out the leaking sheath and replace it with a new unknown sheath.There was no reported patient injury.The operator was trained to the brite tip sheath.The devices were opened in sterile field.The devices were stored, handled, and prepped per the instructions for use (ifu).There was no difficulty experienced in prepping the devices.There were no anomalies noted prior to inserting into the patient.There was no damage noticed to the devices prior to opening the package.There was no difficulty removing the devices from the sterile packaging.The device was not pulled from the packaging by the hub.There were no reported patient injuries.The product was not returned for analysis.A product history record (phr) review of lot 17908009 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿hemostasis valve/hub ¿ separated - during use¿ and ¿ hemostasis valve/hub - leakage - during use¿ could not be confirmed in one of the devices as it was not returned for analysis.The reported ¿hemostasis valve/hub ¿ separated - during use¿ and ¿ hemostasis valve/hub - leakage - during use¿ was not confirmed in the returned device as no damages were observed and during functional flushing test no leakage was noted.The exact cause of the reported event could not be conclusively determined.Based on the information reported and the analysis of the returned device, it is impossible to determine what may have contributed to the reported events.According to the instructions for use which is not intended as a mitigation of risk ¿remove air from csi (including side port extension) by inverting and flushing with heparinized saline or suitable isotonic solution¿.The ifu also warns users to discard the device if there are damages noted during inspection.Neither the phr review nor the product analysis suggests that the reported failure could be related to the manufacturing process of the unit.No corrective or preventive actions will be taken.This is one of two devices involved in the reported event.The reporting reference number of the related device is 9616099-2020-04001.
|
| |
|
Event or Problem Description
|
|
As reported, the valve of the two (2) 7f 23cm brite tip catheter sheath introducers (csi) were leaking and the cap came off the sheath when the user took out the unknown wire from the sheath.The physician was able to take out the leaking sheath and replace it with a new unknown sheath.There was no reported patient injury.The operator was trained to the brite tip sheath.The devices were opened in sterile field.The devices were stored, handled, and prepped per the instructions for use (ifu).There was no difficulty experienced in prepping the devices.There were no anomalies noted prior to inserting into the patient.There was no damage noticed to the devices prior to opening the package.There was no difficulty removing the devices from the sterile packaging.The device was not pulled from the packaging by the hub.There were no reported patient injuries.Only one device will be returned for analysis.
|
| |
|