| Model Number |
M0036125100 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the subject coil was attempted to insert as a framing coil.Subject coiled was pushed, and pulled several times, and prematurely detached within the microcatheter.Physician proceeded to remove the subject coil from microcatheter, and exchanged it with a similar product.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to the reported main coil prematurely detached/separated during use.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the subject coil was attempted to insert as a framing coil.Subject coiled was pushed and pulled several times and prematurely detached within the microcatheter.Physician proceeded to remove the subject coil from microcatheter and exchanged it with a similar product.No clinical consequences were reported to the patient due to this event.
|
| |
|