Model Number RBY4C0835-B
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-01802.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the portal vein using pod coils, ruby coils, and a lantern delivery microcatheter (lantern).It was reported that the patient anatomy was tortuous.During the procedure, the physician injected an absorbable gelatin sponge.While advancing the first pod coil through the mid-shaft of the lantern, the physician experienced resistance.Subsequently, the pod coil was removed, and the physician implanted another pod coil.While advancing the next coil, a ruby coil, through the mid-shaft of the lantern, the physician experienced resistance.Therefore, the ruby coil was also removed.The procedure was completed using two additional pod coils, and the same lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10714267
Report Number3005168196-2020-01803
Device Sequence Number11689645
Product Code HCG
UDI-Device Identifier00814548018386
UDI-Public00814548018386
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/28/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBY4C0835-B
Device Catalogue NumberRBY4C0835
Device Lot NumberF86878
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2020
Initial Report FDA Received Date10/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-