| Model Number |
N/A |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of aseps0010 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported via medwatch "when deaccessing a port-a-cath, the port needle did not retract into the safety.The needle had to be physically pulled out of the port and the safety was not able to be applied." placement info: port a cath single lumen-right internal jugular no harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a cracked infusion set y-site luer was confirmed and the cause appeared to be supplier-related.The product returned for evaluation was two 20ga x 0.75¿ powerloc max safety infusion sets with y-site.The first sample (sample 1) exhibited usage residues throughout.Needleless injection caps were attached to both luer adapters.A longitudinally aligned split was observed in the y-site, extending from the orifice past the attached accessory.Microscopic inspection of sample 1 revealed that the split occurred along a molding weld line.The split surfaces were dull and granular.Sample 2 was received in its original sealed packaging.The sample was unused.Microscopic inspection of sample 2 confirmed a molding weld line in the y-site but was otherwise unremarkable.The crack occurred along a feature caused during the molding process.The device is a supplied component and corrective actions have been implemented by the supplier.The complaint sample was manufactured prior to implementation of those actions.H3 other text : evaluation findings are in section h.11.
|
| |
|
Event or Problem Description
|
|
It was reported via medwatch "when deaccessioning a port-a-cath, the port needle did not retract into the safety.The needle had to be physically pulled out of the port and the safety was not able to be applied." placement info: port a cath single lumen-right internal jugular no harm to patient.
|
| |
|