|
Additional information: h6, h10: device evaluation: one smiths medical cadd administration set was returned for analysis in a used condition.Visual inspection of the pump showed delamination at the join of the filter with the tube in the sample received.During functional testing, a leak was observed fleeing from the air vent of the filter in the sample, the reported complaint was able to be confirmed.Based on the evidence, the complaint was confirmed, and the problem source of the reported event is unknown.
|