Model Number DM3500
Device Problems Data Problem (3196); Intermittent Communication Failure (4038)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, after the loop recorder was placed, an error message was received when attempting to confirm the sensing on the device.The device was repositioned and the error message occurred again.A magnet was placed over it to see if the bluetooth would work, but it was not successful.The device was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of data problem and intermittent communication failure was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10714896
MDR Text Key212451135
Report Number2017865-2020-15924
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/05/2021
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000072040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age76 YR
-
-