• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RD SET NEO-NON-STERILE; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RD SET NEO-NON-STERILE; OXIMETER Back to Search Results
Model Number 4003
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported that the device provided low readings.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.  during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed.  .
 
Event or Problem Description
The customer reported that the device provided low readings.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RD SET NEO-NON-STERILE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key10714947
Report Number3011353843-2020-00172
Device Sequence Number3174315
Product Code DQA
UDI-Device Identifier10843997009598
UDI-Public10843997009598
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Model Number4003
Device Catalogue Number4003
Device Lot Number20DVH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer12/02/2020
Initial Report FDA Received Date10/21/2020
Supplement Report FDA Received Date12/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age6 MO
Patient Weight1
-
-