| Model Number |
PM3210 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that an asymptomatic patient presented to a follow-up in-clinic with noise over-sensing causing inappropriate automatic mode switch episodes from their pacemaker.Noise was able to be reproduced during isometric exercises.Pacemaker mediated tachycardia was also noted.No intervention was performed and patient was stable after the event.
|
| |
|