Model Number 20
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified the reported issue.Physio found broken pieces of the device's therapy pcb assembly inside the bottom case, damage to the j2 capacitor connector, and damage to the front and top cases.The therapy pcb assembly, capacitor connector, front and top cases were replaced to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation by physio-control it was found that the device logged a critical event code.The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.
 
Event or Problem Description
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation by physio-control it was found that the device logged a critical event code.The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.
 
Additional Manufacturer Narrative
Although the customer initially reported their device was functional, physio-control found that the customer's device was unable to deliver energy in the reported condition.Physio-control further evaluated the removed therapy pcb assembly found diodes cr43 and cr44 were shorted.Physio determined that the cause of the reported issue was likely shorted diode cr44.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK 20 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS NON-WEARABLE
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key10715051
Report Number0003015876-2020-01560
Device Sequence Number19816644
Product Code MKJ
UDI-Device Identifier00883873813737
UDI-Public00883873813737
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20
Device Catalogue Number99402-000001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer11/09/2020
Initial Report FDA Received Date10/21/2020
Supplement Report FDA Received Date11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-