Model Number 283512
Device Problem Corroded (1131)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the device was leaking.Per service reports, this complaint can be confirmed.It was observed during evaluation that the device was leaking.Further, the motor was found to be corroded and not turning smoothly.The keypad assembly was not responding properly.Moreover, the resistance values of the hand control and the protective earth resistance of the cable were out of range.The motor, motor cable, and hand control set were replaced to resolve the issues.The device was cleaned, tested and found to be fully functional.Fluid ingress into the device and contact with the motor is responsible for the rusty motor.Corroded motor has a tendency to stick and not turn.The drifting resistance value would have caused the keypad assembly not to respond properly.With the available information, we cannot determine the root cause of the other identified failures.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 09/04/2017 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via phone that the device leaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10715690
MDR Text Key212549900
Report Number1221934-2020-03148
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other
Type of Report Initial
Report Date 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Date Manufacturer Received10/09/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-