| Model Number |
MMT-723NAP |
| Medical Device Problem Code |
Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Pump received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test due to blank display.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was exposed to moisture.Customer states they do not hear fluid when shaking the pump.Customer stated they did not see fluid under the lcd.Customer stated the pump was not dropped.Customer stated there were no cracks in the pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|