• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PROLENE BLU 30IN 3-0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROLENE BLU 30IN 3-0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Model Number 8528H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: what do you mean by "the sutures which break at a low tension threshold" do you mean the sutures broke during the procedure? yes, the sutures broke without exerting excessive tension.Did 6 wires break during this single patient's procedure? yes : (b)(6) 2020 during the installation of a heartware.The following information was requested, but unavailable: what tissue was being approximated or sutured when it broke? events submitted via (b)(4).
 
Event or Problem Description
It was reported that the patient underwent a cardiac surgery on (b)(6) 2020 and the suture was used.During installation of heart ware, a fragility of suture which break at a low tension threshold was noted.The suture broke without exerting excessive tension.Another like suture box from the same batch was opened to finish the procedure without problem.There were no clinical consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 11/18/2020.Additional information: d4, d10, h4, h6.A manufacturing record evaluation was performed for the finished device lot number qbbehd/8528h56, and no non-conformances related to the reported complaint condition were identified.Additional h3 investigation summary: an unopened sample of product code 8528, lot # qbbehd were received for evaluation.During the visual inspection of the unopened sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The device performed without any defect noted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE BLU 30IN 3-0 D/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key10717236
Report Number2210968-2020-08193
Device Sequence Number3060851
Product Code GAW
UDI-Device Identifier10705031091108
UDI-Public10705031091108
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number8528H
Device Catalogue Number8528H
Device Lot NumberQBBEHD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer10/26/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date11/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-