• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETLW1624C156E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Occlusion (1984); Obstruction/Occlusion (2422)
Date of Event 10/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in an unknown endovascular treatment.It was reported that an angiogram performed over three years later, showed that the left limb etlw1624c156e had a slight narrowing in the proximal portion and moderate to severe thrombus throughout the entire left limb.Intervention was performed on the same date, the entire limb was relined down to the hypogastric artery on the left side.This resolved the event.As per the physician the cause of the event could not be determined.No additional clinical sequelae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Additional information received; it was confirmed the thrombus extended from the bifurcate down through to the limb.Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: etbf3216c145e s/n:(b)(6) ; use by date: 27-jun-2019; upn 00613994990983.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10717761
Report Number9612164-2020-04055
Device Sequence Number11574760
Product Code MIH
UDI-Device Identifier00643169467491
UDI-Public00643169467491
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/30/2019
Device Model NumberETLW1624C156E
Device Catalogue NumberETLW1624C156E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/01/2020
Supplement Date Received by Manufacturer10/22/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date11/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-