Catalog Number PFSR200027
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/05/2020
Event Type  malfunction  
Event Description
It was reported that before a navio tka procedure, after initially calibrating a "re-initializing camera" message was prompted when the system said the camera became disconnected.They checked cable connections and after several minutes of waiting for hardware connection to reestablish without success, they proceeded to reboot the system in addition to re-connecting the cables.After multiple attempts of these steps without successful camera connection, they resorted to swapping the camera for the backup camera the hospital had in storage.The cases proceeded without further error messages.The delay was less than 30 minutes and no patient was harmed.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the ndi camera polaris spectra, part number pfsr200027, serial (b)(6) and intended to be used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NDI CAMERA POLARIS SPECTRA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10718006
MDR Text Key212492537
Report Number3010266064-2020-01934
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556649541
UDI-Public885556649541
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR200027
Was Device Available for Evaluation? No
Date Manufacturer Received10/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CABLE, CAMERA CART, NAVIO/200028.; NAVIO SURGICAL SYSTEM US/NPFS02000. SERIAL:(B)(6)
-
-