• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 32 MM DIAMETER; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 32 MM DIAMETER; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Limited Mobility Of The Implanted Joint (2671)
Date of Event 09/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2020-00270, and 3007963827-2020-00271.Medical devices: femur cemented cruciate retaining (cr) standard right size 11 catalog#: 42502607002 lot#: 64420164, articular surface medial congruent (mc) right 12 mm height catalog#: 42522100912 lot#: 64262616, tibia cemented 5 degree stemmed right size h catalog#: 42532008302 lot#: 64446188.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the patient underwent right knee manipulation under anesthesia approximately three months post-implantation due to stiffness.No additional patient consequences or revisions have been reported.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: b4, g4, g7, h2, h3, h6, and h10.Reported event was confirmed by review of medical records.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: 1 month follow up: full weight bearing with normal gait and no pain, normal alignment, rom 10-90°, health score 95, neutral-dissatisfied.At the 3 month follow up: full weight bearing with normal gait and no pain, normal alignment, rom 10-86°, health score 90, neutral-satisfied.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL POLY PATELLA CEMENTED 32 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key10718469
Report Number0002648920-2020-00468
Device Sequence Number16909335
Product Code JWH
UDI-Device Identifier00889024247635
UDI-Public(01)00889024247635(17)280229(10)64620955
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42540000032
Device Lot Number64620955
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer12/02/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date12/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight94
-
-