(b)(4).Batch # p56x3v.Investigation summary: the analysis results found that the ats45nk device (b) was received with the firing mechanism damaged and with three tr45w loads received.The reloads (c, d & e) were received partially fired 1/10 and the reload (d) with the reload lock out spring damaged.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the firing trigger teeth was found broken.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.It should be noted that as part of our quality process, each device is visually inspected, and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information will be included in complaint trending that is reviewed on a regular basis to determine if further action is necessary.Reload c: tr45w p55806 partially fired 1/10; reload d: tr45w p55806 partially fired 1/10, spring damaged; reload e: tr45w, p5570g partially fired 1/10.
|
|
It was reported that during the distal radical gastrectomy for gastric cancer surgery, could not fire when severing small intestine tissue on the 1st firing.Changed another ats45nk and tr45w to continue the surgery, the same problem happened again.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|