• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SMALL 3.5MM DEPTH GAUGE SHORT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SMALL 3.5MM DEPTH GAUGE SHORT; PLATE, FIXATION, BONE Back to Search Results
Model Number 71175069
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the evos small 3.5 mm depth gauge short is broken.It is unknown when this malfunction occurred and if there was any patient involvement.No injuries reported.
 
Additional Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the tip is broken off the evos small 3.5mm depth gauge short.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 3.5MM DEPTH GAUGE SHORT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10720474
Report Number1020279-2020-05653
Device Sequence Number3501100
Product Code HRS
UDI-Device Identifier00885556635872
UDI-Public00885556635872
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175069
Device Catalogue Number71175069
Device Lot Number18FRS0016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2020
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date11/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-