SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 3 8MM; TEMPLATE Back to Search Results
Model Number 00584102308
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/29/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during tka procedure the articular surface provisional sz 3 8mm cracked in half while being used to trial the implants inside the patient.All pieces recovered.There was a s&n back up device.No delay.No injuries reported at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned instrument confirms the stated failure mode.The articular surface provisional has fractured into two pieces.Both pieces were returned for evaluation.This instrument exhibits signs of significant use and wear.The articular provisional was manufactured in 2014.A medical investigation was conducted and this case reports that the trial cracked in half while trailing the implants.Per email communication, all pieces were recovered from the patient, without injury or delay.The procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 3 8MM
Common Device Name
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10721858
Report Number1020279-2020-05665
Device Sequence Number19708413
Product Code HWT
UDI-Device Identifier00885556619445
UDI-Public00885556619445
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00584102308
Device Catalogue Number00584102308
Device Lot Number62696620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer11/17/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date12/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-